Measurement Specialties Inc. Model 9116 User’s Manual
Page 89 www.MEAS-SPEC.com
RE-BOOT MODULE (UDP/IP Command ‘psireboot’)
Purpose: To unconditionally “reboot” a specified module.
Command
“psireboot ethadr”
where ethadr is the Ethernet address of the specified module in the
following special hex-digit format
‘xx-xx-xx-xx-xx-xx’
Response
none (module reboots)
Description: When a Model 9116 module receives this broadcast command, (by continuously
monitoring port 7000) it responds by immediately restarting its firmware. The
result is essentially the same as a power-up restart, in that any TCP/IP
connection is lost, and the module returns to its normal startup state. The host
must wait long enough for the re-boot process to be completed before it can
again request a connection to the module.
Example:
Re-boot a specified module on the network.
“psireboot 00-E0-8D-00-00-01”
Response: