Measurement Specialties Inc. Model 9116 User’s Manual
Page 111 www.MEAS-SPEC.com
Figure 5.5
Solenoids in Module
Figure 5.5
Solenoids in Module
(4) Ensure the three (3) solenoid manifold O-rings are present and free of contaminants.
Replace the solenoid with the new one by carefully aligning and gently tightening the
screws.
(5) Attach the wiring harness to the solenoid and connector P6 of the PC-322 Main Board
Assembly.
(6) Reassemble the module.
(7) Test your scanner to enusre proper operation.
5.1.6 Replacement of O-Rings
MEAS calibration valves include static and dynamic O-ring seals. When used properly, the
rated durability of the dynamic O-rings is in excess of 1,000,000 shifts of the calibration valve.
The procedures described below should be used for replacement of all the O-rings in the Model
9116 Intelligent Pressure Scanner. Use the tools and follow the general warnings already
described at the start of Section 5.1.
The material needed for the O-ring replacement can be acquired through the proper
maintenance kit available from MEAS. Specifically needed for these procedures are calibration
manifold and piston O-rings, PTFE cup seals, a fast evaporating cleaning fluid ( i.e. acetone,
alcohol, Freon
®
, etc.), 50 psi dry air supply, and Krytox
®
fluorinated grease (read product
warnings and recommendations thoroughly). Service of O-ring seals requires a clean
working environment. Introduction of contaminants to the O-ring or internal calibration
manifold surfaces can result in internal pneumatic leaks. Inspection for contaminates generally
requires some type of magnification device, such as a microscope.